Processing DAU/FPA Tuning data
RG Bias Sweeps
The RG Bias sweep is performed by stepping each RG Mag voltage invididually and collecting multiple raw captures at each voltage level.
Pre-test Setup:
- Open 2 instances of the DDC Raw Viewer. One set to CCD ID Blue and the other CCD ID Red.
It's important that there are only 2 instances open, otherwise they may overwrite each others exported data.
As the sweeps are performed, the raw captures will be automatically exported to $HOME/ocidata/raw_captures
, where $HOME
is the user's home directory. The sub-directory names and structures will be generated from the XINA markers. The data for a single RG Mag sweep will be placed into its own directory e.g. 2022-04-13T00-04-58_RgM3_Red
. This directory will contain the raw captures for each voltage step.
The scripts to process the data should be run on this directory. For example:
cd ~/ocidata/raw_captures
analyze_dau_fpa 2022-04-13T16-09-49_RgM3_Red_Dark
For light sweeps, you need to provide the analyzed results from the dark sweep.
analyze_dau_fpa 2022-04-13T00-04-58_RgM3_Red 2022-04-13T16-09-49_RgM3_Red_Dark/analysis