Alerts
Alerts are messages generated during mining that often serve as a notice or warning that a certain threshold or event occurred. Each alert contains a severity, a time, and a specific message.
MOMA Alerts
General Error Check
- Alert ID: 1
- Severity: Notice
- Description: Grep message log for "Error"
MAIF APV Comm
- Alert ID: 2
- Severity: Warning
- Description: Grep message log for "ldi: error"
Spacewire Comm Error
- Alert ID: 3
- Severity: Warning/Notice
- Description: Spacewire Parity (HK 522) > 0 or Spacewire Disconnect (HK 523) > 0 after 30sec
MOMA Safe Event
- Alert ID: 4
- Severity: Notice
- Description: HK 507 MAX, MOMA_STATUS_SAFE > 0 at any time during a TID
WRP Too Fast
- Alert ID: 5
- Severity: Notice
- Description: PWM Speed - HK 832 MAX > 103000
WRP Failed To Start
- Alert ID: 6
- Severity: Warning
- Description: PWM Speed - HK 832, MAX > 1000 and < 99000 RPM
WRP Pressure Spike
- Alert ID: 7
- Severity: Warning
- Description: HK 830, MOT_SUP_IMON, MAX > 1.1A
Dark Counts
- Alert ID: 8
- Severity: Warning
- Description: HK 252 SCAN_TELEM_DAC_ID = 20000 AND HK 253 SCAN_TELEM_DAC_VAL > 5
Laser Burst Energy Variation
- Alert ID: 9
- Severity: Info
- Description: Std. Dev. of burst pulse energy is > 5.0
Laser Pulses Mismatch
- Alert ID: 10
- Severity: Notice
- Description: Main Pulses != # of Pulses in Burst
Unknown SEB sequences
- Alert ID: 11
- Severity: Notice
- Description: Scan type is LDI or EI, and SEB packet sequence count == 0
Filament VMON Pegged
- Alert ID: 12
- Severity: Warning
- Description: Anytime Filament VMON HK is > 4.9
Expected Pulse Count Mismatch
- Alert ID: 13
- Severity: Warning
- Description: HKID 1050 (PLS_SUBPKT_CNT) != (pkt.length() - 4 / 16)
Gap in Scans
- Alert ID: 14
- Severity: Warning
- Description: Bin Delta between Science packets is > 100
Old Packet Format
- Alert ID: 15
- Severity: Info
- Description: Scan Packet Version Number != 2 (c_scan.science_data_pkt_version())
Missing Scan Status Packets
- Alert ID: 16
- Severity: Warning
- Description: Number of Scan Status Packets != Number of Scans in Summed Packet | Disable for Dark Counts
Dropped Packets
- Alert ID: 17
- Severity: Warning
- Description: Dropped packets are detected by gaps in the packets' sequence numbers.
Both EMs On
- Alert ID: 18
- Severity: Warning
- Description: Both Electron Multipliers are on at the same time
Sub-system Communication Errors
- Alert ID: 19
- Severity: Warning
- Description: Communication errors between any of the sub-systems (MAIF, SEB, GC, LASER) as reported by the CDH/FSW.