MOMA Alerts A MOMA Alert is defined as a notification of certain parameters of interest in a single MOMA Test. This page details the logic we use to extract the data for each alert. Note that this is the master reference, which means the engineers and scientists should ensure that the logic detailed here is correct. All alert data is in an Alert Panel within a typical TID record __TOC__ General Error Check Severity : Notice Message Log is parsed for the phrase 'Error' while excluding messages that contained 'Holding at stop for operator' or 'error =' <- Means a value was being set. MAIF APV Comm Check Severity : Warning Message Log is parsed for the phrase 'ldi: error'. Spacewire Comm Error Severity : Warning Spacewire Disconnect (HK 522) > 0 or Spacewire Parity (HK 523) > 0 after 30 sec. Severity : Notice Spacewire Disconnect (HK 522) > 0 or Spacewire Parity (HK 523) > 0 only before 30 sec. MOMA Safe Event Severity : Notice MOMA_STATUS_SAFE (HK 507) is 1 at any time during the TID. WRP Too Fast Severity : Notice PWM Speed (HK 832) > 103000RPM at any time during the TID. WRP Failed to Start Severity : Warning PWM Speed (HK 832) surpassed 1000RPM but never reached over 99000RPM at any time during the TID. WRP Pressure Spike Severity : Warning MOT_SUP_IMON (HK 830) > 1.1A at any time during the TID. Dark Counts Severity : Warning SCAN_TELEM_DAC_ID (HK 252) == 20000 and SCAN_TELEM_DAC_VAL (HK 253) > 5 at any time during the TID. Laser Burst Energy Variation Severity : Info Standard Deviation of the Energy from all Pulses in a Single Burst > 5.0 Laser Pulses Mismatch Severity : Notice Number of Main Pulses Commanded (HK 1002) != # of Pulses in the Burst Unknown SEB Sequences Severity : Notice Scan Type is LDI or EI && SEB Packet Sequence Count == 0 Scan Mode determined using c_scan.scan_mode_as_string() Sequence Length determined using len(c_scan.seb_sequence_pkt_indexes()) Filament VMON Pegged Severity : Warning FIL_VMON (HK 19) > 4.9 at any time during the TID. Pulse Count | Packet Size Mismatch Severity : Warning HKID 1050 (PLS_SUBPKT_CNT) != (pkt.length() - 4 / 16) Summed Scan | TIC Mismatch Severity : Warning In a sum scan, the number of summed scans != the number of TICs Gap in Scans Severity : Warning Bin Delta between Science packets is > 100 cFactor | TIC Mismatch Severity : Warning Number of cFactors != # of TICs Scan Status | TIC Mismatch Severity : Warning Number of Scan Status Packets != # of TICs Dropped Packets Severity : Warning Packet Sequence Numbers Both EMs On Severity : Warning Both Electron Multipliers are on at the same time Limit Violations Severity : Warning Come Back Later